DETERMINACION DE As, Fe, Sb, Ca Y Si EN FANGOS ARSENICALES POR FLUORESCENCIA DE RAYOS X

Authors

  • J. Jiménez Chacón Instituto de Materiales y Reactivos para la Electrónica, Universidad de la Habana
  • A. Boza Carbonell Instituto de Materiales y Reactivos para la Electrónica, Universidad de la Habana
  • M. Bustamante Sánchez Instituto de Materiales y Reactivos para la Electrónica, Universidad de la Habana
  • R. Fernández Reyes Instituto de Materiales y Reactivos para la Electrónica, Universidad de la Habana

Abstract

An X-Ray Fluorescence analysis procedure was developed for the determination of As, Fe, Sb, Ca and Si in arsenical
sludge. The samples were prepared as a mixture with soluble starch in a 4:1 ratio; they were pressed at 20 MPa for
30 s on a H3BO3 backing, obtaining tablets of about 40 mm in diameter, resistant to handling. In the X-ray excitation
source, tubes with W and Rh anodes were used, operated at 40 kV and 40 mA. The measurements were carried out in
vacuum. To carry out the calibration, synthetic standards were used and the inter-elemental effects were corrected
using the Lucas-Tooth and Pyn.e model. In addition, the selection criteria for interfering elements proposed by
Plesch was used and it was found that for As, Fe, Sh and Si, it is not necessary to consider all the elements in the
sample during calibration.

Published

2022-11-23

How to Cite

Jiménez Chacón, J., Boza Carbonell, A. ., Bustamante Sánchez, M. ., & Fernández Reyes, R. . (2022). DETERMINACION DE As, Fe, Sb, Ca Y Si EN FANGOS ARSENICALES POR FLUORESCENCIA DE RAYOS X . NATIONAL CENTER FOR SCIENTIFIC RESEARCH (CENIC) CHEMICAL SCIENCES JOURNAL, 24(1-2-3), 047-050. Retrieved from https://revista.cnic.cu/index.php/RevQuim/article/view/3103